SUPPORT

Typical Problems - Noviplex

Email Support / Request Free Sample

Partial wetting of the plasma collection disc

Possible Causes

  • card was bent

  • insufficient blood applied

  • top layer was disturbed prior to blood collection

Solutions

  • keep cards isolated prior to use

Blood in unexpected areas of the card

Possible Causes

  • top layer was opened prematurely

  • card was flexed excessively

Solutions

  • leave Noviplex card in pouch until use to prevent problems

*For first time customers only at research universities in the US.